National Kidney Disease Education Program National Kidney Disease Education Program National Kidney Disease Education Program National Kidney Disease Education Program National Kidney Disease Education Program National Kidney Disease Education Program
   
Información en Español
NKDEP Home
 
 
Patients & Public
Health Professionals
Laboratory Professionals
Recommendations
Resources
FAQ
Publications
Contact Us
Resources
News & Events
About NKDEP

SEARCH NKDEP:
  
Home > Resources > Commutability Study Results
Laboratory Professionals

Commutability Study Results

NIST SRM 967 and CAP LN24 — Spring/Summer 2006

Reference materials are typically used to establish or validate the traceability of a routine measurement procedure to a value assigned to the material by a reference measurement procedure of higher order. When a reference material is intended to be measured by a routine clinical method, commutability must be validated among all of the routine methods for which the reference material is intended for use. Commutability has been defined as the equivalence of the mathematical relationships between the results of different measurement procedures for a reference material and for representative samples from healthy and diseased individuals.1 This definition is based on the more technical definition in ISO document 17511.2 Commutability does not imply accuracy or trueness of results, only that results for a reference material had the same mathematical relationship between two methods that was observed for native clinical samples measured by the same two methods.

Fresh frozen serum creatinine reference materials from the National Institute of Standards and Technology (NIST) Standard Reference Material (SRM) 967 and from the College of American Pathologists’ (CAP) Accuracy Calibration Verification/Linearity Survey LN24 were found to be commutable with a panel of native clinical serum samples for the manufacturers’ reagents and instruments listed below. Reference materials and clinical samples were measured by the stated analytical system and by a LC/IDMS reference measurement procedure at NIST. Commutability was evaluated using criteria described in Clinical and Laboratory Standards Institute (CLSI) document EP14-A2.3

Commutable Reagents and Instruments

Manufacturer Reagent (Product Number) Instrument Model
Abbott Diagnostics Creatinine(7D64) Abbott ARCHITECT cSystemsTM c8000
Abbott Diagnostics Creatinine (7D64) Abbott AEROSET® System Abbott Diagnostics Creatinine, Next Generation (3L81) Abbott ARCHITECT cSystemsTM c8000
Abbott Diagnostics Creatinine, Next Generation (3L81) Abbott AEROSET® System
Siemens Diagnostics Creatinine_2 (B01-4126-01) ADVIA® 1650 Chemistry
Siemens Diagnostics Creatinine Enzymatic (B01-4127-01) ADVIA® 1650 Chemistry
Beckman Coulter, Inc. CREA (442760) Synchron® CX4 PRO
Beckman Coulter, Inc. CREA (442760) Synchron® Unicel DxC 800
Beckman Coulter, Inc. Creatinine (443340) Synchron® CX3 Clinical System
Beckman Coulter, Inc. CREm (472525) Synchron® Unicel DxC 800
Dade Behring, Inc. Dimension® CREA Flex (DF33A) Dimension® RxL System
Ortho Clinical Diagnostics CREA Slides (814 1947) VITROS® 950 Chemistry System
Olympus Diagnostics Systems Creatinine (OSR6178) Olympus AU400®
Olympus Diagnostics Systems Creatinine (OSR6178) Olympus AU640e®
Roche Diagnostic Corporation CREA plus (11775642) Hitachi 911 Analyzer
Roche Diagnostic Corporation CREA/RB (12217333) Hitachi 911 Analyzer

The list provided here documents commutability of the NIST SRM 967 and CAP LN24 reference materials with native clinical serum samples for various reagent-instrument combinations determined for specific reagents used on these specific instrument systems and may not apply to different reagent-instrument combinations. However, since these reference materials met the CLSI EP14-A2 criteria for commutability with all reagent-instrument combinations tested, one would expect these reference materials to be commutable with native clinical serum samples when the same reagent is used on very similar instruments using the same methodological principle. Nevertheless, caution should be exercised in extrapolating the above findings to different reagent-instrument combinations, even when the instruments and reagents are all produced by the same IVD manufacturer. It is recommended that clinical laboratories consult with the manufacturer(s) of their reagent and instrument to determine the applicability of the established commutability characteristics for these reference materials to particular configurations of creatinine reagents and instrument systems.

References

  1. Miller WG, Myers GL, Rej R. Why commutability matters. Clin Chem 2006;52:553-4.
  2. ISO International Standard 17511:2003(E). In vitro diagnostic medical devices - Measurement of quantities in biological samples - Metrological traceability of values assigned to calibrators and control materials. First edition, 2003-08-15. International Organization for Standards, Geneva, Switzerland, 2003.
  3. CLSI Document EP14-A2. Evaluation of Matrix Effects; Approved Guideline – Second Edition. Clinical and Laboratory Standards Institute, Wayne, PA, USA, 2005.

 


Back to Top

Site Map | Disclaimer | Privacy | Accessibility | Contact Us

Last Reviewed: February 22, 2007

NKDEP is an initiative of the National Institute of Diabetes and Digestive and Kidney Diseases (NIDDK),
National Institutes of Health (NIH), U.S. Department of Health & Human Services (DHHS).

U.S. Department of Health and Human Services logo National Institutes of Health logo National Institute of Diabetes and Digestive and Kidney Diseases Logo USA.gov logo